Fields of application

JUMO tecLine H₂O₂ and PAA are membrane amperometric sensors for measuring the concentration of hydrogen peroxide or peracetic acid in liquids. They are used in the food, beverage, dairy, pharmaceutical and chemical industries, as well as in water disinfection and washing processes. Analog and digital versions are available for various measuring ranges.

 

Features:

  • Hydrogen peroxide (H₂O₂) or peracetic acid (PAA) measurement
  • Membrane amperometric technology with 2 electrodes
  • Measuring ranges up to 200,000 mg/l, depending on version
  • Analog 4-20 mA or digital RS485 Modbus RTU output
  • Automatic temperature compensation
  • According to the version – resistance to chemicals and surfactants
  • Easy calibration using a connected controller
  • SPECIFICATIONS
  • AVAILABLE MODELS
  • DOCUMENTS

SPECIFICATIONS

  • Device type

    Membrane amperometric sensor
  • Measured parameter

    Hydrogen peroxide (H₂O₂) or peracetic acid (PAA), depending on the version
  • Measuring principle

    Amperometric, 2 electrodes
  • PAA measurement ranges

    Analog: 0–200; 0–500; 0–2000; 0–5000; 0–20,000 mg/l.

    Digital: 0–200; 0–2000; 0–20,000 mg/l
  • H₂O₂ measurement ranges

    Analog: 0–500; 0–2000; 0–10,000; 0–20,000; 0–50,000 mg/l.

    Digital: 0-20,000; 0-200,000 mg/l
  • Available sensor versions

    H₂O₂ or PAA; analog 4-20 mA or digital RS485
  • Membrane type

    Chemical-resistant membrane; resistance to surfactants depends on version
  • Response time (t90)

    PAA: approx. 3 min. (analog), approx. 5 min. at +10 °C (digital). H₂O₂: approx. 5-10 min. (analog), approx. 8 min. (digital)
  • Operating water temperature

    0 to +45°C;
  • Ambient temperature

    0 to +55°C
  • Working pH range

    H₂O₂: pH 2-11; PAA: pH 1-6
  • Pressure resistance

    Up to 1 bar relative pressure (2 bar absolute); pressure fluctuations are not allowed
  • Temperature compensation

    Automatic, integrated temperature sensor
  • Required water flow

    Approx. 15 cm/s (30 l/h in JUMO flow fitting)
  • Stabilization time

    H₂O₂: approx. 3 hours; PAA: approx. 1 hour during first start-up
  • Housing material

    PVC-U, stainless steel 1.4571
  • Sensor dimensions

    Ø25 mm; length approx. 220 mm (analog) or 205 mm (digital)
  • Power supply

    12-30 V DC (analog); 22.5-26 V DC (digital)
  • Analog output

    4-20mA
  • Digital link

    RS485, Modbus RTU
  • Electrical connection

    2-pin terminals (analog) or 5-pin M12 (digital)
  • Calibration

    According to the concentration determined by the laboratory method using the connected controller
  • Maintenance

    Electrolyte replacement every 3-6 months; membrane heads – approximately once a year
  • Available mounting options

    Individual or combined flow fittings, mounting brackets, flow control equipment
  • Additional equipment

    Replaceable membrane heads, electrolyte, connecting cables, JUMO AQUIS and dTRANS controllers

AVAILABLE MODELS

DOCUMENTS

  • Brochure: Liquid Analysis

    10.32 MB

    Download
  • Brochure: Product range – (overview in DIN A4 format)

    5.62 MB

    Download
  • Brochure: Water and Wastewater

    9.87 MB

    Download
  • Brochure: JUMO digiLine – the intelligent, bus-compatible digital sensor connection system for liquid analysis

    5.24 MB

    Download
  • Brochure: Product range

    9.72 MB

    Download
  • Operating instructions: JUMO tecLine H2O2 / JUMO tecLine PAA Hydrogen peroxide or peracetic acid sensor, type 202636

    1.6 MB

    Download
  • Technical Guide: Measurement of Hydrogen Peroxide and Peracetic Acid

    1.75MB

    Download
  • Data sheet: JUMO tecLine H2O2 / JUMO tecLine PAA, Hydrogen peroxide and peracetic acid sensor – 202636

    483.85 KB

    Download
  • Certificate: China RoHS-e – JUMO tecLine H2O2 and PAA

    96.69 KB

    Download
  • Order information: 202636

    113.33 KB

    Download
  • UK Declaration of Conformity: JUMO tecLine H2O2 and PAA

    724.33 KB

    Download
  • Interface description: Modbus – JUMO tecLine Cl, TC, ClO2+O3, H2O2+PAA, Br

    347.82 KB

    Download