Measurement of waste combustion exhaust gases from boilers and incinerators or gases from various industrial furnaces.
Measurable components
NO, SO2, CO2, CO, CH4, O2
Measurement principles
Non-dispersive infrared absorption method.Single source and beam - NO, SO2, CO2, CO, CH4
Electrochemical or paramagnetic method (measuring element incorporated in theanalyser) or zirconia method - O₂
Repeatability
±0.5% of full scale
Linearity
1% of full scale
Before correction for atmospheric pressure (optional)
Zero drift
±2% of full scale/week
Auto zero calibration required for
scale ≤ 500 ppm.
Scale drift
±2% of full scale/week
Response time (for 90% of full scale)
Electrical response: 1 to 15 sec
Between 10 and 30 sec including cell purge time. This time depends on the number of gases measured and their scales.
Power supply
Voltage: 100 V to 240 V AC
Scale range: 85 V to 264 V AC
Frequency: 50Hz/60Hz
Operating temperature
-5°C to 45°C (40°C for 2 optical benches at 200 VA AC)
Operating humidity
Max. 90% RH (non-condensing)
Housing
Steel for indoor mounting
Color
Front panel; light gray (PANTON 1C-F)
Display
4-digit digital display (LED backlit LCD)
Dimensions
133 × 483 x 382 mm (HxWxD);
(19" rack mounting)
Weight
11 kg
Product code | Price |
---|---|
ZPA |
Certificate : TUV QAL1 infrared gas analyser ZPA
251.51 KB
DownloadCertificate: UKCA analyser infrared gas ZPA
78.7 KB
DownloadCertificate: EU declaration of conformity analyser infrared gas ZPA
70.46 KB
DownloadManual: analyser infrared gas ZPA
3.44 MB
DownloadManual: analyser infrared gas ZPA Modbus communication
561.63 KB
DownloadDatasheet: Infrared gas analyser ZPA
1.84 MB
DownloadBrochure: analyser infrared gas ZPA reliability and precision
4.55 MB
DownloadBrochure: Gas analysers, reliable solutions for gas analysis
8.47 MB
DownloadBrochure: Measures for energy and the environment
4.01 MB
Download